Skip to main content
Advertisement
  • Loading metrics

Sensitive, Noninvasive Detection of Lymph Node Metastases

Abstract

Background

Many primary malignancies spread via lymphatic dissemination, and accurate staging therefore still relies on surgical exploration. The purpose of this study was to explore the possibility of semiautomated noninvasive nodal cancer staging using a nanoparticle-enhanced lymphotropic magnetic resonance imaging (LMRI) technique.

Methods and Findings

We measured magnetic tissue parameters of cancer metastases and normal unmatched lymph nodes by noninvasive LMRI using a learning dataset consisting of 97 histologically proven nodes. We then prospectively tested the accuracy of these parameters against 216 histologically validated lymph nodes from 34 patients with primary cancers, in semiautomated fashion. We found unique magnetic tissue parameters that accurately distinguished metastatic from normal nodes with an overall sensitivity of 98% and specificity of 92%. The parameters could be applied to datasets in a semiautomated fashion and be used for three-dimensional reconstruction of complete nodal anatomy for different primary cancers.

Conclusion

These results suggest for the first time the feasibility of semiautomated nodal cancer staging by noninvasive imaging.

Introduction

Most primary malignancies spread systemically via lymphatic dissemination [1]. For example, the finding of axillary nodal metastases predicts a much shorter disease-free survival in breast cancer [2]. The total nodal tumor burden (number of affected nodes and metastatic tumor volume) affects prognosis even more severely [3]. Accurate lymph node staging also remains a cornerstone in choosing the most appropriate therapy for a given stage. Therapeutic intervention of metastatic lymph nodes [4], prophylactic radiation of frequently affected drainage routes [5], and systemic therapies [6] all have been shown to improve survival. Genetic profiles identifying metastatic tumors [7], serum biomarkers, and proteomic profiles are currently being developed to identify patients at risk [8,9]. No direct genetic profile, however, has been demonstrated to date to accurately predict the presence of human nodal metastases in a given patient. Rather, surgical approaches, such as sentinel lymph node biopsy or lymph node dissection, are still commonly used. Careful histological analysis includes mapping, bisectioning, and rapid staining in the frozen tissue laboratory. Higher diagnostic accuracies can be achieved by serial sectioning (50 μm) and by immunohistochemical staining [10,11].

Noninvasive imaging studies are commonly used during the workup of primary malignancies. Typically, lymph nodes are diagnosed by tomographic techniques (computed tomography [CT], magnetic resonance imaging [MRI]) as malignant when their short axis is >10 mm in size [12]. Such size criteria, however, have been shown to be unreliable [13]. Similarly, the detection of cancer in nonenlarged (occult) nodes is often quite low by positron-emission tomography (PET) and single photon emission computed tomography imaging. For example, small nodal metastases (< 5 mm) are often missed by PET imaging in patients with breast cancer [14]. More recently, it has become possible to image anatomic regions at submillimeter resolutions by MRI, with excellent spatial coverage and reduced motion artifacts. The development [15,16] and clinical introduction of lymphotropic magnetic nanoparticles has been shown to significantly improve diagnostic accuracies of MRI for nodal staging (LMRI) in prostate cancer [17]. These nanoparticles serve as probes for lymphatic anatomy and function and enhance tumor detection through abnormal distribution patterns in malignant nodes [17,18].

Despite the advances of LMRI for cancer staging, image analysis has been challenging and occasionally controversial. Traditional analysis has been based on a reader's identification of certain structural abnormalities that can be variable, given differences in acquisition parameters and interpretation criteria [19,20,21]. Furthermore, it has been challenging to quickly and accurately analyze large datasets generated by LMRI.

The goal of the current study was to develop and test technologies that would vastly improve the accuracy of current LMRI nodal staging. Specifically we set out to (a) determine whether unique magnetic parameters existed and could be used for semiautomated image analysis and (b) whether the technique could be applied to different primary cancers. Here we provide the first comprehensive analysis of tissue parameters validated against histopathology as an end point.

Methods

Study Design

The Institutional Review Board approved the current study and all patients signed informed consent. The study was divided into a learning (n = 97 lymph nodes with known histopathology) and a test dataset (n = 216 lymph nodes with known histopathology; Table 1). Assignment into datasets was done in temporal fashion. The learning dataset represented retrospective cases at outset of the study, and the test dataset represented prospective cases collected during a 1-y interval. In the learning set, 55% of the nodes were benign, and 45% of the nodes were malignant. The learning dataset was obtained from 36 patients (24 male, 12 female, age 28–85 y, mean 59.7 y) with histologically proven primary genitourinary malignancies (prostate, 21; bladder, 9; testes, 5; ureter, 1). All patients completed the MRI study and then underwent surgical resection (n = 26) and/or nodal biopsy (n = 10). The investigated nodes had a mean short axis diameter of 10.5 mm (range 3–39 mm).

The test dataset was obtained from 34 patients (25 male, nine female, age 30–82 y, mean 58.9 y) with histologically proven malignancies from different primaries (Table 1), including prostate (n = 18), breast (n = 7), penile (n = 4), bladder (n = 2), testes (n = 2), and colon (n = 1). Seventy-nine percent of the nodes were benign and 21% of the nodes were malignant. The nodes in the test dataset had a mean short axis diameter of 10.0 mm (range 3–39 mm). Both datasets included the full spectrum of normal nodes to completely replaced nodes.

MRI

MRI was performed at 1.5 T (System 9X, General Electric Medical Systems, Milwaukee, Wisconsin, United States) using phased-array coils. All images were archived on DICOM PACS servers (MIPortal, CMIR and Siemens Medical Systems, Erlangen, Germany; and Impax RS 3000, AGFA Technical Imaging Systems, Richfield Park, New Jersey, United States) for subsequent analysis. Images of the pelvis (n = 56) extended from the pubic symphysis to just above the level of aortic bifurcation. In patients with primary testicular cancers (n = 7) imaging was extended superiorly to include the renal hilum and retroperitoneum. In patients with breast cancer (n = 7) we obtained MR images of the bilateral axillae, including the internal mammary and supraclavicular regions. All patients were imaged with identical pulse sequences and timing parameters. Imaging was performed before and 24 h after intravenous ferumoxtran-10 administration (Combidex, Advanced Magnetics, Cambridge, Massachusetts, United States; 2.6 mg Fe/kg diluted in normal saline and infused over a 20-min period using a 5-μm filter).

The acquired pulse sequences included (a) axial T2-weighted fast spin-echo (TR/TE, 4500/80; flip angle, 90°; field of view, 24–28 cm; slice thickness, 3 mm; matrix, 256 × 256; number of excitations, 2; in-plane resolution, 1.2 mm); (b) a T1-weighted two-dimensional gradient-echo sequence obtained in different anatomical planes (TR/TE 175/1.8; flip angle, 80°; field of view, 22–30 cm; slice thickness, 4 mm; matrix, 128 × 256; in-plane resolution, 2.0 mm); (c) an axial T2-weighted dual TE gradient-echo (TR/TE 2100/14–24; flip angle, 70°; field of view, 26–28 cm; slice thickness, 3 mm; matrix, 160 × 256; in-plane resolution, 1.7 mm); and (d) a three-dimensional (3D) T1-weighted gradient echo sequence; TR/TE 4.5–5.5/1.4; flip angle, 15°; field of view, 24–28 cm; slice thickness, 1.4 mm; matrix, 256 × 256; in-plane resolution, 1.0 mm).

The above listed imaging sequences and parameters had previously been optimized to reduce motion artifacts, maximize signal-to-noise ratio (SNR), and provide diagnostically useful images of the pelvis, abdomen, and chest within clinically acceptable time limits. The T2-weighted fast spin-echo sequence, in (a) above, was primarily used for qualitative nodal detection, and hence a square pixel with more than one acquisition was obtained. The two-dimensional axial T1-weighted gradient-echo sequence, in (b) above, was chosen to achieve adequate anatomical coverage within a short imaging time. The axial dual-echo gradient-echo sequence, in (c) above, was developed specifically for this project to provide artifact-free datasets for quantitative image analysis. A matrix size of 160 × 256 was chosen for this sequence to achieve a balance between the upper limits for imaging time while reducing image noise. Finally, a 3D T1-weighted sequence was obtained, in (d) above to provide a dataset for vascular maximum intensity projection (MIP) reconstructions.

Quantitative Image Analysis

All image analysis was performed on archived DICOM images using different software packages (e.g., custom-built software such as CMIR-Image, MGH, Boston, Massachusetts, United States; Syngo, Siemens Medical Systems; Advantage Windows, General Electric Medical Systems). Lymph nodes were identified by readers who manually placed kernels onto each node for automated boundary detection and calculation of nodal dimensions and volumes. The thus identified regions of interest (ROIs) encompassed the entire lymph node (not only portions of it) and were used for quantitative signal-intensity (SI) measurements (see Table 2). Serial measurements of nodal dimensions on different pulse sequences or time points varied less than 2%.

thumbnail
Table 2. Frequency of Imaging Parameters in Learning Dataset

https://doi.org/10.1371/journal.pmed.0010066.t002

A number of quantitative tissue parameters were calculated either as differences between pre- and postcontrast scans (δ) or as single-value analysis on postcontrast scans (see Table 2). The lymph node/muscle (LNM) ratio was calculated by dividing signal intensities of an entire lymph node by that of adjacent muscle using a similar-sized ROI, drawn manually. The nodal SI change was calculated by obtaining SI before and after contrast administration. The nodal SNR was calculated by obtaining SD/SDnoise. The T2* was calculated in nodal ROIs on dual TE images using CMIR-Image. T2* maps were constructed by performing fits of a standard exponential relaxation model (S = Ke–TE/T2*) to the data on a pixel-by-pixel basis. Only pixels with intensity greater than a threshold level (2X of noise) were considered during the fitting process. Pixel variance was obtained from post-MR images. Comparative visual analysis included short axis measurements, and identification of heterogeneity, large focal defects, and central hyperintensity, according to criteria previously established [12,17].

To determine the diagnostic accuracy of the different tissue parameters in the learning dataset, we determined sensitivity, specificity, and predictive values for each parameter alone and in combination (Table 3). The most discriminatory parameters were then applied to the test dataset (Table 4).

thumbnail
Table 3. Discriminatory Power of Imaging Parameters in Learning Dataset

https://doi.org/10.1371/journal.pmed.0010066.t003

thumbnail
Table 4. Application of Quantitative Parameters to Test Dataset (n = 216)

https://doi.org/10.1371/journal.pmed.0010066.t004

In the final set of semiautomated image analysis, 3D reconstructions were obtained for nodal mapping onto vascular anatomy using MIP projections. While the MIP projections do not aid in the differentiation between malignant and benign lymph nodes, they are invaluable in providing anatomic content to the dozens of lymph nodes identified. In particular, MIP images were generated interactively from postcontrast, fat-saturated, volumetric interpolated breath-hold images to outline vascular anatomy. The evaluated lymph nodes characterized as benign or malignant (by T2*/variance analysis) were then superimposed on the volumetric 3D images, using customized software (Advantage Windows, General Electric Medical Systems).

Statistical Analysis

Data were expressed as mean ± standard deviations (SD) and medians. All statistical testing was performed using GraphPad Prism (GraphPad Software, San Diego, California, United States). The significance between two individual groups was determined using the nonpaired Student's t-test (e.g., benign and malignant datasets in Figure 1). For the more discriminatory datasets alternative-free-response receiver operating characteristic curves were plotted. Ratios for cut-off single-value parameters were defined to yield highest sensitivity and specificity. Accuracy for a given parameter was expressed as the area under the curve (Az), and values are summarized in Table 4.

thumbnail
Figure 1. Tissue Parameters in Learning Dataset

Nodal tissue parameters for benign and malignant nodes are shown before (A and B) and after (C–E) intravenous administration of magnetic nanoparticles. Note the insensitivity of conventional MRI (A and B), better separation using single-value analysis (C and D) and excellent separation using two-value analysis (E).

https://doi.org/10.1371/journal.pmed.0010066.g001

Histology

All lymph nodes were sampled histologically within 2 wk of the MRI (mean: 6 d; range: 2–14 d). The analysis was done in surgically resected lymph nodes (n = 55; both benign and malignant nodes) or in fine needle aspirates and core biopsies (n = 15; malignant nodes only), implementing careful mapping procedures to correlate nodes. Surgically excised nodes were sectioned at 10–20 μm intervals after bihalving and were stained with hematoxylin-eosin.

Results

Learning Dataset

The learning dataset consisted of 97 histologically validated lymph nodes from 36 patients with different primary malignancies (see Table 1). The mean short axis diameter was 10.5 mm (range 3–39 mm) with 56 of the 97 nodes (58.3%) measuring less than 10 mm, that is, below the traditional imaging cutoff for malignancy (“occult nodes”). Table 2 summarizes the incidence of different visual, comparative (before and after contrast administration), and semiautomated (postcontrast administration only) parameters in the two different groups. Figure 1 is a graphical representation of overlaps between malignant and benign groups for different parameters listed in Table 2. Table 3 summarizes sensitivities, specificities, and predictive values for the different quantitative imaging parameters. Sensitivities of metastasis detection by visual image analysis ranged from 50%–94%, however, often with lower specificities. Volumetric measurements, in particular, were insensitive markers of malignancy in nonenlarged nodes (see Table 3).

In contradistinction, image analysis of pre- and postcontrast image sequences resulted in higher specificities and sensitivities (see Table 3). Comparative differences between benign and malignant nodal groups were highest for T2* and pixel variance measurements (see Table 3). Of all the semiautomated parameters tested alone, T2* measurements showed the highest sensitivity (93%; 95% confidence interval: 82%–98%) and specificity (94%; 95% confidence interval: 84%–99%) in the learning dataset (see Figure 1 and Table 3).

Of all the semiautomated parameters tested in combination, T2* measurements combined with pixel variance analyses postcontrast showed the highest sensitivity (98%; 95% confidence interval: 88%–99%) and specificity (94%; 95% confidence interval: 82%–98%) in the learning dataset (Figure 1E). Using the dual-value analysis, there was one malignant outlier in the benign dataset (the lymph node was 3 mm in overall size, with few malignant cells seen on histology, and probably too small for analysis) and two benign outliers in the malignant dataset (both these nodes showed hyalinosis replacing more than 50% of the nodal architecture).

Test Dataset

To determine whether feature extraction would be accurate for prospective nodal staging, we utilized the above criteria against a larger test dataset encompassing 216 validated lymph nodes from 34 patients, including different primaries (see Table 1). The sensitivity, specificity, and predictive values of the most discriminatory parameters of this prospective analysis are summarized in Table 4. We primarily focused on semiautomated image analysis of postcontrast scans because of the high sensitivity and specificity determined in the learning dataset. T2* measurements showed a sensitivity of (93%; 95% confidence interval: 82%–99%) and a specificity of (91%; 95% confidence interval: 85%–96%). Combined T2* and pixel variance analysis achieved a sensitivity of 98% (95% confidence interval: 88%–99%) and a specificity of 92% (95% confidence interval: 87%–96%) comparable to that of the learning set and much superior to currently used size criteria.

Using the dual-value analysis, there were two malignant outliers in the benign dataset (both of these nodes were less than 3 mm in overall size and probably too small for analysis—similar to the learning dataset) and three benign outliers in the malignant dataset (two of these nodes had hyalinosis replacing more than 50% of the nodal architecture and one had macrocalcifications). More important, all the misclassified nodes occurred in individual patients rather than in the same patient and, hence, did not affect the overall nodal staging on a patient-by-patient basis in this dataset.

Image Reconstruction

Utilizing semiautomated feature extraction to identify lymph nodes and image analysis (based on T2* and pixel variance), we subsequently proceeded to map individual lymph nodes onto vascular anatomy in the different anatomic drainage patterns. Figure 2 summarizes the different steps in image analysis. Figure 3 and Video S1 shows an example of a 45-y-old patient with colorectal cancer undergoing semiautomated nodal staging. In this particular patient, MRI identified six positive lymph nodes (< 10 mm each), reconstructed as a 3D dataset, whereas all positive lymph nodes were missed by PET scans. Figure 4 and Video S2 show reconstructions and analyses from a patient with a breast cancer primary with bilateral nodal metastases. Note the high spatial resolution allowing the detection of a 3-mm nodal metastasis.

thumbnail
Figure 2. Steps in Semiautomated Image Analysis

Semiautomated image analysis involves recognition and automated segmentation of each lymph node (A), quantitation of magnetic tissue parameters (T2*, variance of pixel values; [B]), comparison of extracted tissue parameter to a database (C), and 3D reconstruction of nodal anatomy onto vascular anatomy (D).

https://doi.org/10.1371/journal.pmed.0010066.g002

thumbnail
Figure 3. Pelvic Nodal Staging

Nodal staging in patient with colorectal cancer. A PET scan using 18FDG as a tracer (A) and a CT scan (B) were interpreted as negative for nodal metastases. LMRI identified six small pelvic lymph nodes ([C] and [D]; red arrowheads), which had magnetic parameters of malignancy. Semiautomated reconstruction (E) identifies multisegmental metastases, subsequently proven histologically (F). For 3D reconstruction of pelvic nodal anatomy see Video S1.

https://doi.org/10.1371/journal.pmed.0010066.g003

thumbnail
Figure 4. Breast Cancer Mapping

Patient with breast cancer prior to sentinel lymph node biopsy.

(A) Conventional axillary MRI shows nonenlarged lymph nodes that do not meet the size criteria of malignancy (white bar = 5 mm).

(B) Following intravenous administration of nanoparticles, a single 3-mm intranodal metastasis was correctly identified.

(C) Ex vivo MRI of sentinel node specimen confirms metastasis.

(D) Semiautomated nodal analysis and reconstruction correctly juxtaposed solitary lymph node metastases adjacent to two normal lymph nodes.

(E) Correlative histopathology confirms the diagnosis. For 3D reconstruction of axillary nodal anatomy see Video S2.

https://doi.org/10.1371/journal.pmed.0010066.g004

Discussion

We show that it is feasible to extract various quantitative tissue parameters to predict the likelihood of nodal metastases in vivo. These results are highly relevant in cancer staging because they provide evidence that (a) quantitative tissue parameters enable diagnosis of lymph node metastases while reducing interobserver variability and (b) that semiautomated reconstructions allow spatially more extensive mapping than is currently possible.

Metastases to lymph nodes occur during growth of most primary malignancies, and their presence mandates the need for more extensive and systemic therapy. Nodal cancer staging currently relies on invasive procedures (surgical lymph node dissection, sentinel lymph node resection, biopsy) with significant morbidity and cost [22,23], or insensitive tomographic imaging methods [24]. For example, detection sensitivities using size criteria with state-of-the-art multislice CT are as low as 50%, whereas PET imaging of nonenlarged nodes has equally low sensitivities [14]. Based on the observation that nanoparticulate solutions accumulate in nodal macrophages upon systemic injections [25,26], lymphotropic superparamagnetic preparations have been developed [16]. In earlier clinical trials (using lower spatial resolution sequences), metastases of 1–2 mm have been detected [17], whereas as few as 1,000 tumor cells have been detected in nodes in experimental mouse models [18]. Despite these advances, it has been difficult to acquire images of sufficiently high resolution and to derive parameters to automate diagnosis. The data presented here indicate that unique magnetic parameters allow identification of nodal metastases and accurate 3D reconstructions, including surgically inaccessible lymph nodes.

The significance of the above findings is 3-fold. First, the ability to directly and noninvasively monitor nodal tumor involvement represents a powerful diagnostic tool for cancer. Accurate staging represents the cornerstone for triaging patients to either localized or to more aggressive and systemic therapies. Second, the method described here was sensitive for the limited subsets of primary cancers tested. It is reasonable to hypothesize that such analysis could be applied to staging of other common primaries. In particular, lung, colorectal, genitourinary, and head and neck cancers could benefit from this staging procedure. In addition to nodal staging, the nanoparticle-enhanced MRI can also be used to measure microvascularity in primary tumors [27] and to improve the detection of liver metastases [28]. Third, our results are significant because the semiautomated staging method is highly accurate and reduces variability in visual image analyses between different observers.

The LMRI staging technique is believed to be clinically relevant in several key areas. First, LMRI may play a significant role in avoiding unnecessary surgeries, that is, those in node-positive patients. Second, since LMRI can detect lymph nodes outside traditional surgical fields, this information may influence surgical approaches. In colorectal cancer, LMRI may provide a “sentinel-node-like” guide to staging. Third, it is likely that LMRI would be useful to identify appropriate patients to receive neoadjuvant chemotherapy prior to surgery. Currently, neoadjuvant therapy is often reserved for postoperative patients, once the nodal status has been determined. Fourth, LMRI may be particularly useful to guide radiation therapy by mapping the complete nodal status onto bony and vascular landmarks. Finally, LMRI could be used to avoid invasive diagnostic procedures, which are not part of therapy. For example, LMRI could replace lymphangiography, mediastinoscopy, or endoscopic ultrasound for nodal staging.

Our findings have a number of direct implications for technology development and in clinical care. Accurate measurements of T2* relies on motion artifact-free multiecho pulse sequences that are not routinely available on clinical scanners at spatial resolutions required for nodal staging. Such sequences will have to be implemented and combined with postprocessing tools to simplify and semiautomate analysis. Similar software approaches are already used routinely in lung nodule characterization [29] or screening for breast cancers [30]. We predict that in the case of LMRI, such automation routines will be highly specific, given the unique mechanism of image contrast. As a proof-of-principle, we implemented approaches to identify, segment, analyze, and display nodal information. While the current technology is already highly accurate, we anticipate further improvements with hardware and software advances. We hope that this will ultimately translate into clinical practice and replace unnecessary intervention.

Supporting Information

Video S1. Automated 3D Reconstruction of Pelvic Nodal Anatomy

https://doi.org/10.1371/journal.pmed.0010066.s001

(7.5 MB MPEG)

Video S2. Automated 3D Reconstruction of Axillary Nodal Anatomy

https://doi.org/10.1371/journal.pmed.0010066.s002

(5.1 MB MPEG)

Acknowledgments

The authors would like to acknowledge Drs. M. Saksena and M. Torabi for help with data collection, Dr. Elkan Halpern for statistical advice, Drs. Guimaraes and Ross for critical review and Benjamin King for assistance in early data analysis. We would also like to acknoweledge Drs. P. F. Hahn, S. Gazelle, and R. Seethamraju for many helpful discussions. Parts of the study were funded by National Institutes of Health grants P50 CA86355 and NO1-CM037008, a Center for Molecular Imaging Research development grant, a grant to Massachusetts General Hospital from Siemens Medical Systems, and a General Electric Medical Systems–Association of University Radiologists fellowship grant (MH). Combidex was provided by Advanced Magnetics. The funding sources had no role in study design, collection, analysis, and interpretation of data, in the writing of the report, or in the decision to submit the paper for publication.

Author Contributions

MGH conducted the study, analyzed the data, and enrolled patients. RW designed and conducted the study and analyzed the data. RW and MGH contributed to writing the paper.

References

  1. 1. Hanahan D, Weinberg RA (2000) The hallmarks of cancer. Cell 100: 57–70.
  2. 2. Cummings MC, Walsh MD, Hohn BG, Bennett IC, Wright RG, et al. (2002) Occult axillary lymph node metastases in breast cancer do matter: Results of 10-year survival analysis. Am J Surg Pathol 26: 1286–1295.
  3. 3. Kell MR, Winter DC, O'Sullivan GC, Shanahan F, Redmond HP (2000) Biological behaviour and clinical implications of micrometastases. Br J Surg 87: 1629–1639.
  4. 4. Sikov WM (2000) Locally advanced breast cancer. Curr Treat Options Oncol 1: 228–238.
  5. 5. Diab SG, Hilsenbeck SG, de Moor C, Clark GM, Osborne CK, et al. (1998) Radiation therapy and survival in breast cancer patients with 10 or more positive axillary lymph nodes treated with mastectomy. J Clin Oncol 16: 1655–1660.
  6. 6. Henderson IC, Berry DA, Demetri GD, Cirrincione CT, Goldstein LJ, et al. (2003) Improved outcomes from adding sequential Paclitaxel but not from escalating Doxorubicin dose in an adjuvant chemotherapy regimen for patients with node-positive primary breast cancer. J Clin Oncol 21: 976–983.
  7. 7. Ramaswamy S, Ross KN, Lander ES, Golub TR (2003) A molecular signature of metastasis in primary solid tumors. Nat Genet 33: 49–54.
  8. 8. Sidransky D (1997) Nucleic acid-based methods for the detection of cancer. Science 278: 1054–1059.
  9. 9. Wulfkuhle JD, Liotta LA, Petricoin EF (2003) Proteomic applications for the early detection of cancer. Nat Rev Cancer 3: 267–275.
  10. 10. Pargaonkar AS, Beissner RS, Snyder S, Speights VO (2003) Evaluation of immunohistochemistry and multiple-level sectioning in sentinel lymph nodes from patients with breast cancer. Arch Pathol Lab Med 127: 701–705.
  11. 11. Matsuda J, Kitagawa Y, Fujii H, Mukai M, Dan K, et al. (2004) Significance of metastasis detected by molecular techniques in sentinel nodes of patients with gastrointestinal cancer. Ann Surg Oncol 11: 250S–254S.
  12. 12. Jager GJ, Barentsz JO, Oosterhof GO, Witjes JA, Ruijs SJ (1996) Pelvic adenopathy in prostatic and urinary bladder carcinoma: MR imaging with a three-dimensional TI-weighted magnetization-prepared-rapid gradient-echo sequence. AJR Am J Roentgenol 167: 1503–1507.
  13. 13. Anzai Y, Piccoli CW, Outwater EK, Stanford W, Bluemke DA, et al. (2003) Evaluation of neck and body metastases to nodes with ferumoxtran 10-enhanced MR imaging: Phase III safety and efficacy study. Radiology 228: 777–788.
  14. 14. Guller U, Nitzsche E, Moch H, Zuber M (2003) Is positron emission tomography an accurate non-invasive alternative to sentinel lymph node biopsy in breast cancer patients? J Natl Cancer Inst 95: 1040–1043.
  15. 15. Weissleder R, Elizondo G, Wittenberg J, Rabito CA, Bengele HH, et al. (1990) Ultrasmall superparamagnetic iron oxide: Characterization of a new class of contrast agents for MR imaging. Radiology 175: 489–493.
  16. 16. Weissleder R, Elizondo G, Wittenberg J, Lee AS, Josephson L, et al. (1990) Ultrasmall superparamagnetic iron oxide: An intravenous contrast agent for assessing lymph nodes with MR imaging. Radiology 175: 494–498.
  17. 17. Harisinghani MG, Barentsz J, Hahn PF, Deserno WM, Tabatabaei S, et al. (2003) Noninvasive detection of clinically occult lymph-node metastases in prostate cancer. N Engl J Med 348: 2491–2499.
  18. 18. Wunderbaldinger P, Josephson L, Bremer C, Moore A, Weissleder R (2002) Detection of lymph node metastases by contrast-enhanced MRI in an experimental model. Magn Reson Med 47: 292–297.
  19. 19. Bellin MF, Lebleu L, Meric JB (2003) Evaluation of retroperitoneal and pelvic lymph node metastases with MRI and MR lymphangiography. Abdom Imaging 28: 155–163.
  20. 20. Harisinghani MG, Saini S, Slater GJ, Schnall MD, Rifkin MD (1997) MR imaging of pelvic lymph nodes in primary pelvic carcinoma with ultrasmall superparamagnetic iron oxide (Combidex): Preliminary observations. J Magn Reson Imaging 7: 161–163.
  21. 21. Holland AE, Hendrick RE, Jin H, Russ PD, Barentsz JO, et al. (2000) Correlation of high-resolution breast MR imaging with histopathology: Validation of a technique. J Magn Reson Imaging 11: 601–606.
  22. 22. Blanchard DK, Donohue JH, Reynolds C, Grant CS (2003) Relapse and morbidity in patients undergoing sentinel lymph node biopsy alone or with axillary dissection for breast cancer. Arch Surg 138: 482–487. discussion, 487-488.
  23. 23. Campbell SC, Klein EA, Levin HS, Piedmonte MR (1995) Open pelvic lymph node dissection for prostate cancer: A reassessment. Urology 46: 352–355.
  24. 24. Stell DA, Carter CR, Stewart I, Anderson JR (1996) Prospective comparison of laparoscopy, ultrasonography and computed tomography in the staging of gastric cancer. Br J Surg 83: 1260–1262.
  25. 25. Dumont AE, Martelli A (1969) X-ray opacification of hepatic lymph nodes following intravenous injection of tantalum dust. Lymphology 2: 91–95.
  26. 26. Dumont AE, Martelli AB, Schinella RA (1982) The intranodal distribution of lymph-borne particles injected intravenously. Br J Exp Pathol 63: 479–484.
  27. 27. Bremer C, Mustafa M, Bogdanov A, Ntziachristos V, Petrovsky A, et al. (2003) Steady-state blood volume measurements in experimental tumors with different angiogenic burdens: A study in mice. Radiology 226: 214–220.
  28. 28. Harisinghani MG, Saini S, Weissleder R, Halpern EF, Schima W, et al. (1997) Differentiation of liver hemangiomas from metastases and hepatocellular carcinoma at MR imaging enhanced with blood-pool contrast agent Code-7227. Radiology 202: 687–691.
  29. 29. Henschke CI, McCauley DI, Yankelevitz DF, Naidich DP, McGuinness G, et al. (1999) Early Lung Cancer Action Project: Overall design and findings from baseline screening. Lancet 354: 99–105.
  30. 30. Burnside ES, Rubin DL, Shachter RD, Sohlich RE, Sickles EA (2004) A probabilistic expert system that provides automated mammographic-histologic correlation: Initial experience. AJR Am J Roentgenol 182: 481–488.

Patient Summary

When deciding on treatment for patients with cancer, it is very important to assess whether the cancer has spread to lymph nodes—both to help decide what treatment a patient should have and what the eventual outcome might be. Previous ways of finding involved lymph nodes included taking out the nodes by surgery, ultrasound, and CT and MRI scans.

A solution of magnetic nanoparticles that tend to go to lymphoid organs was injected and then tracked by MRI. The pattern of the particles was abnormal when there was metastasis in the nodes, and it was possible to train a computer to recognize this abnormality. The authors developed the program in one group of patients and then tested it in another group, in which they were able to correctly predict whether the nodes were involved in about nine of ten nodes. In addition, they could use the information to display a virtual picture of the involved nodes.

The technique will need to be validated in a larger group of patients, and by other investigators. However, it means that it is potentially possible to work out much more precisely, and with less chance of error, whether lymph nodes are involved in cancer. Hence, treatment can be better planned, and if surgery is needed to remove nodes for analysis, then this technique could ensure that the surgery is as minimal as possible.

RadiologyInfo, a public information site developed by the American College of Radiology and the Radiological Society of North America: http://www.radiologyinfo.org/

Medline Plus, which has health information from the National Library of Medicine: http://www.nlm.nih.gov/medlineplus/cancer.html